dc.description.abstractMoja bakalárska práca sa zaoberá témou Stabilita nastavenia centra rotácie u kamery InfiniaJa som sa vo svojej práci podrobne oboznámil so scintilačnou kamerou Infinia Hawkeye, ktorá je umiestnená na oddelení nukleárnej medicíny v Českých Budějoviciach. Cieľom mojej práce bolo zistiť, či centrum rotácie (C.O.R.) prebieha okolo pevnej osy a teda či nedochádza k jeho pohybu. Metóda ktorú som zvolil pre splnenie mojich cieľov bolo testovanie kamery, ktoré sa vykonávalo 360° akvizíciou okolo bodového zdroja 99mTc. Bodový zdroj bol prichytený na tyč a umiestnený na pacientovom lôžku kamery Infinia. Testovanie bolo prevedené s dvoma kolimátormi LEHR a LEGP v režimoch H a L. Toto testovanie vykonával rádiologický fyzik, vždy popoludní aby sa zabezpečil plynulý chod pracoviska. Testom sa potvrdila moja hypotéza, že hodnoty sa budú pohybovať v rozmedzí hodnôt udávaných výrobcom. Testovanie sa vykonávalo raz mesačne po dobu pol roka. Práve preto, že dnes vnímame obrovský technický pokrok, a vznikajú prístroje, ktoré nám umožňujú vytvárať čoraz kvalitnejšie a detailnejšie obrazy a prispievajú k lepšej diagnostike a liečbe je veľmi dôležité dbať na kvalitu kontroly. Pokroky v konštrukcii gama kamier sa za posledných niekoľko rokov zlepšili vo všetkých aspektoch kvality obrazu, hlavne pri tomografických obrazoch. Keďže zložitosť systému sa zvyšuje, začína byť dôležité, aby všetci pracovníci nukleárnej medicíny boli schopní rozoznať rôzne typy odchýlok, ktoré sa môžu vyskytnúť v zobrazovacích systémoch nukleárnej medicíny. Primárnym účelom programu kontroly kvality na oddeleniach nukleárnej medicíny je overiť, že získané obrazy budú presne odrážať rozloženie radiofarmaka vo vnútri pacienta. Úlohou kontroly kvality je odhaliť zmeny v gama kamerovom systéme, ktoré môžu nepriaznivo ovplyvňovať interpretáciu klinických štúdií. Existuje veľké množstvo faktorov, ktoré sa podieľajú na konečnej kvalite obrazu ( či už je to homogenita, časové či energetické rozlíšenie, frekvencia impulzov, systém stredu rotácie, rotačná stabilita hlavy detektora). Počas každodenných vyšetrení je obmedzené množstvo času, ktoré môže byť rozumne venované systému kontroly kvality. Z toho dôvodu hlavným cieľom programu kontroly kvality by malo byť sledovanie tých parametrov, ktoré sú najviac citlivé na zmeny vo výkone systému a s najväčšou pravdepodobnosťou majú vplyv na klinickú štúdiu. Špeciálna séria skúšok kontroly kvality bola navrhnutá tak, aby sa zabezpečil optimálny výkon kamery. Po celú dobu klinickej prevádzky kamery, technici alebo fyzici v nemocnici pravidelne vykonávajú rôzne testy kontroly kvality, niektoré denne, iné týždenné alebo mesačne. Účelom týchto testov je sledovať deň za dňom výkon kamery a odhaliť zmeny z východiskovej hodnoty. Dodatočné testy kontroly kvality môžu byť vykonané na porovnanie správania sa rozličných systémov a optimalizáciu kvality klinických štúdii.cze
dc.format74
dc.format74
dc.language.isocze
dc.publisherJihočeská univerzitacze
dc.rightsBez omezení
dc.subjectnukleárna medicínacze
dc.subjectscintilačná kameracze
dc.subjectcentrum rotáciecze
dc.subjectbodový zdrojcze
dc.subjectkontrola kvalitycze
dc.subjectnuclear medicineeng
dc.subjectscintillation cameraeng
dc.subjectcentre of rotationeng
dc.subjectpoint sourceeng
dc.subjectquality controleng
dc.titleStabilita nastavení centra rotace u kamery Infiniacze
dc.title.alternativeCentre of rotation stability setting for the INFINIA cameraeng
dc.typebakalářská prácecze
dc.identifier.stag38826
dc.description.abstract-translatedMy bachelor thesis focus on Centre of rotation stability for Infinia cameraAt my work I got closely familiarize with scintillation camera Infinia Hawkeye, which is located in the department of nuclear medicine in the České Budějovice. The aim of my work was to determine whether the centre of rotation (C.O.R.) takes place around a fixed axis, and thus whether there is not movement. The method that I chose to accomplish my goals, was to test the camera , which was performing by 360° acquisition around a point source 99mTc. The point source was fitted to a rod and placed on the Infinia camera on patient's pallet. Testing was carried out with two collimators LEHR and LEGP in modes H and L. The testing was performed by radiological physicist, always in the afternoon to ensure smooth running of the workplace. The test confirmed my hypothesis that the value will be within the values specified by the manufacturer. Testing was performed once a month for half a year. In In this days because we are apperceiving tremendous technological progress, and devices which allow us to produce more perfectly and detailed images and contribute to better diagnosis and treatment are created it is very important to ensure the quality control. Over the last few years progress in gamma camera designs , have improved all aspects of image quality, particularly in tomographic images. As the system complexity increases, it becomes more important that all nuclear medicine employees are able to recognize the various types of aberrations or artefacts that can occur in nuclear medicine imaging systems. The primary purpose of the quality control program in department for nuclear medicine is to verify that the images accurately reflect distribution of the radiopharmaceutical within the patient. The purpose of quality control is detect changes of gamma camera system that may adversely affect the interpretation of clinical studies. There are large number of factors that contribute to the final quality of the image ( including uniformity, resolution - both intrinsic and energy, pulse frequency, centre of rotation system, rotational stability of the detector head). On a everyday examinations there is a limited amount of time that can be reasonably devoted to quality control system. Hence the main goal of quality control program should be to monitor those parameters that are most sensitive to changes in system performance and are most likely to impact clinical studies. Therefore , the main objective of quality control program should be monitor the parameters that are most sensitive to changes in system performance and most likely have a impact on the clinical studies. Specials series of quality control tests were designed to ensure the optimal performance of the camera. Over the whole time of the entire period of clinical operation, technicians or physicists regularly perform various quality control tests, some daily, others weekly or monthly. The purpose of these tests is monitor day to day camera performance and detect changes from the baselines values. Additional quality control test may be performed to compare characteristics of different systems and to optimize the quality of clinical studies.eng
dc.date.accepted2015-09-09
dc.description.departmentZdravotně sociální fakultacze
dc.thesis.degree-disciplineRadiologický asistentcze
dc.thesis.degree-grantorJihočeská univerzita. Zdravotně sociální fakultacze
dc.thesis.degree-nameBc.
dc.thesis.degree-programSpecializace ve zdravotnictvícze
dc.description.gradeDokončená práce s úspěšnou obhajoboucze
dc.contributor.refereePětivlasová, Eva


Soubory tohoto záznamu

Thumbnail
Thumbnail
Thumbnail
Thumbnail

Tento záznam se objevuje v

Zobrazit minimální záznam